Parameter
|
Range
|
Best Uncertainty
|
|
DC Volts Source
|
1 µV to 1100 V
1.018
10.0 V1100 V to 60 KV
|
4.4 ppm @ 10 VDC
0.7 ppm
0.5 ppm
0.12 % @ 1100 VDC
|
|
DC Volts Measure
|
1 µV to 1 V
1.0 to 1000 V
1.1 kV to 120 KV
|
0.8 ppm @ 1 VDC
0.6 ppm @ 100 VDC0.12 % @ 1.1 kVDC
|
|
DC Current Source
|
100 pA to 100 A
0 to 550A (Clamp-ons)
|
23 ppm @ 100 mADC
0.14 % @ 2A
|
|
DC Current Measure
|
100 pA to 100 A
100 A to 300 A
|
22 ppm @ 100 mADC 0.05 % @ 100 ADC
|
|
AC Voltage Source
|
10 Hz to 1 MHz
2 mV to 22 V
|
20 ppm @ 2 V/1 kHz
|
| |
10 Hz to 300 kHz
22 V to 60 V
|
28 ppm @ 60 V/1 kHz
|
| |
10 Hz to 100 kHz 60 V to 1000 V
|
37 ppm @ 1000V/1 kHz
|
| |
1 kV to 21.2 kV @ 60 Hz
|
0.59 % @ 2 kV/60 Hz
|
| |
21.2 kV to 60 kV @ 60 Hz
|
2.3 % @ 30 kV/60 Hz
|
|
AC Voltage Measure
|
10 Hz to 1 MHz
2 mV to 22 V
|
14 ppm @ 2 V/1 kHz
|
| |
10 Hz to 300 kHz 22 V to 60 V
|
24 ppm @ 60 V/ 1 kHz
|
| |
10 Hz to 100 kHz 60 V to 1000 V
|
32 ppm @ 1000V/1 kHz
|
| |
1 KV to 21.2 KV @ 60 Hz
|
0.58 % @ 2000 V/60 Hz
|
|
AC Current Source
|
1 mA to 11A 10 Hz to 10 kHz
|
86 ppm @ 100 mA/ 1 kH
|
| |
10 µA to 1 mA 10 Hz to 1 kHz
|
160 ppm @ 1 mA/ 1 kHz
|
| |
11 A to 100 A 45 Hz to 440 Hz
|
0.29 % @ 100 A/ 1 kHz
|
| |
0 to 550A(clamp-on only)
|
0.14 % @ 2 A/1 kHz
|
|
AC Current Measure
|
1 mA to 5A
10 Hz to 100 kHz
|
79 ppm @ 100 mA/1 kHz
|
| |
5 A to 20 A
10 Hz to 50 kHz
|
0.10 % @ 10 A/ 1 kHz
|
| |
10 µA to 1 mA 10 Hz to 50 kHz
|
0.09 % @ 1 mA/1 kHz
|
|
DC Resistance Source
|
1 mOHM to 1 MOHM
10.0 MOHM
100.0 MOHM
100 M to 1 GOHM
|
0.21 ppm @ 1 Ohm
5.6 ppm
8.6 ppm
0.12 % @ 1 Gohm
|
|
DC Resistance Measure
|
0.0001 W to 1 GOHM
1 GOHM to 100 TOHM
|
0.26 ppm @ 1 Ohm
0.06 % @ 1 Gohm
|
|
AC Resistance Source
|
0.01 OHM to 100 kOHM@ 1 kHz to 13 MHz
|
0.02 % @ 10 kOhm/1 kHz
|
|
AC Resistance Measure
|
0.1mOHM to 100 OHM
10 Hz to 500 kHz
|
580 ppm @ 100 Ohm /1 kHz
|
|
Capacitance Measure
|
0.1 pF to 1 µF50 Hz to 10 kHz
|
0.011 % @ 1000 pF/1 kHz
|
| |
1 µF to 1 mF10 Hz to 500 kHz
|
580 ppm @ 1 µF/1 kHz
|
|
Capacitance Source
(Individual Values)
|
1000 pF
10 Hz to 1 kHz
|
7.1 ppm @ 1 kHz
|
| |
1 pF to 10 mF
|
0.02% @ 1 kHz
|
|
Inductance Measure
|
100 µH to 10 H10 Hz to 1 MHz
|
0.02% @ 100 mH
|
|
Inductance Source
|
100 µH to 10 H100 Hz to 1 Khz
|
23 ppm @ 100 mH/1 kHz
|
|
RF Power Measure
|
-127 to + 10 dBm
0.1 to 50 GHz
|
2.9 % @ 0 dBm/500 MHz
|
| |
-70 to + 20 dBm
10 MHz to 50 GHz
|
2.9 % @ 0 dBm/1 GHz
|
| |
0.1 to 10 MHz
10 µW to 25 mW
|
0.32 % @ 0 dBm / 10 MHz
|
| |
0.01 to 18 GHz
10 µW to 25 mW
|
0.32 % @ 0 dBm / 1 MHz
|
| |
25 to 1000 MHz
0.3 to 1000 Watts
|
3.5 % @ 50 W/300 MHz
|
|
RF Power Source
|
-127 to + 10 dBm10 MHz to 50 GHz
|
0.7 dB @ 0 dBm/ 1 GHz
|
| |
1 to 100 Watts
0.01 to 1000 MHz
|
3.5 % @ 5 W/300 MHz
|
|
RF Attenuation Measure
|
+10 to -120 dB
50 MHz to 50 GHz
|
0.015 dB @ 0 dBm/ 1 GHz
|
|
RF Attenuation Source
|
10 to 50 dB @ 30 MHz
|
0.008 dB @ 10 dB Attenuation,
0 dBm/ 30 MHz
|
| |
0 to 130 dB
@ DC to 18 GHz
|
0.35 dB @ 2 dB Attenuation,
0 dBm/ 1 GHZ
|
| |
0 to 90 dB
@ 18 to 50 GHz
|
0.93 dB @ 10 dB Attenuation,
0 dBm/20 GHz
|
|
Frequency Source/Measure
|
1 mHz to 50 GHz
|
5 parts in 10-12
|
|
Rise Time Source
|
<200 pS
|
|
|
Rise Time Measure
|
<700 pS
|
pS +/- 3.5%
|
|
Distortion
|
5 Hz to 100 kHz
100 kHz to 600 kHz
|
1.2 % @ 0.1% Distortion
3.5 % @ 1% Distortion
|
|
Modulation Measure
|
AM: 20 Hz to 100 kHz
FM: 20 Hz to 200 kHz
|
1.2% @ 20 MHz/1 kHz\
1.2% @ 20 MHz/400 kHz Dev.
|
|
Modulation Source
|
AM 50 Hz to 100 kHz
FM DC to 200 kHz
|
0.1% @ 10 MHz /1 kHz
0.1% @ 12.5 MHz/5 kHz Dev.
|
|
Phase Source
|
0.000 to 360.0 Degrees
|
6.1 m? @ 0.000 Degrees phaseshift @ 1 V/1 kHz
|
|
Phase Measure
Oscilloscopes
|
0.000 to 360.0 Degrees
50 kHz to 1050 MHz
|
18 m? @ 0.000? shift @ 1 V/1 kHz2.1% @ 50 MHz
|
|
Fiber Optics Source
|
450 to 1700 nm
+10 to ñ90 dBm
|
0.05 dB @ 0 dBm/ 1310 nm
|
| |
632.5, 850, 1310, 1330,
1523, 1550 nm Wavelengths
|
0.003 nm @ 1550 nm
|
|
Fiber Optics Measure
|
450 to 1700 nm
+10 to ñ 110 dBm
|
0.1 dB @ 0 dBm / 850 nm
|
| |
450 to 1700 nm Wavelength
|
0.003 nm @ 1550 nm
|